ACTUAL
SET
Ad¯P -
Calibration of programmer input:
Highest value = 80 °C
ACTUAL
SET
AdP -
Calibration of programmer input:
Lowest value = -20 °C
ACTUAL
SET
Calibration procedure:
In the configuration level, set the programmer input
type to allow control via an external programmer.
0 = Voltage 0 to 10 V
1 = Current 0 to 24 mA
Connect an external programmer to the connector (O).
Set the external programmer to the highest
temperature value.
ACTUAL
SET
In the calibration level, select the item Ad¯P and set
the highest temperature value
(example: 80 °C).
Press the setpoint key (E).
Follow the instructions under section 6.3.1.
page 27
Do not alter any settings on either of the units as long
as the display "Ad¯P" is blinking.
- FC1200-T Recirculating cooler FC1600-T Recirculating cooler FCW2500T Recirculating cooler FCW1600T Recirculating cooler FCW1200T Recirculating cooler FCW1600S Recirculating cooler FCW1200S Recirculating cooler FC1600T Recirculating cooler FC1200T Recirculating cooler FC1600S Recirculating cooler FC1200S Recirculating cooler