5.3.
Bath fluids
Caution:
No liability for use of other bath liquids!
Please contact JULABO before using other than recommended bath fluids.
JULABO takes no responsibility for damages caused by the selection of an
unsuitable bath fluid
Do not use alcohols.
Water:
The quality of water depends on local conditions.
Due to the high concentration of lime, hard water is not suitable for
temperature control because it leads to calcification in the bath.
Ferrous water can cause corrosion - even on stainless steel.
Chloric water can cause pitting corrosion.
Distilled and deionized water is unsuitable. Their special properties cause
corrosion in the bath, even in stainless steel.
Water: - No liablity for use with water.
Danger of freezing at working temperatures <5 °C.
Recommended bath fluids:
Bath fluids
Temperature range
soft/decalcified water
5 °C to 80 °C
See website for list of recommended bath fluids.
Contact: see page 5
5.4. Tubing
Caution:
Employ suitable connecting tubing.
Make sure that the tubing is securely attached.
Avoid sharp bends in the tubing, and maintain a sufficient distance from
surrounding walls.
Regularly check the tubing for material defects (e.g. for cracks).
Preventive maintenance: Replace the tubing from time to time.
Maximum
pressure
Chloroprene
tubing
Textile reinforced tubing
0.5 bar
4.5 bar
- FC1200-T Recirculating cooler FC1600-T Recirculating cooler FCW2500T Recirculating cooler FCW1600T Recirculating cooler FCW1200T Recirculating cooler FCW1600S Recirculating cooler FCW1200S Recirculating cooler FC1600T Recirculating cooler FC1200T Recirculating cooler FC1600S Recirculating cooler FC1200S Recirculating cooler