Operator responsibility – Safety instructions

12


Temperature selection

digital (keypad)

Resolution

°C

0.1

MULTI-DISPLAY indications

LED + LED

Resolution

°C

0.1

Display

accuracy

%

0.5

Temperature stability

°C ±0.2

Temperature control

on/off

Control ratio for feed/return flow temperature, adjustable

% 0 ... 100

Temperature sensor (number)

PTC (3)

Level indication

spy-glass

Error message indication

LED

Electrical connections:
Computer interface

RS 232

Stand-by input

conforming to Namur recommendations

Alarm output

potential-free

Return flow safety device

V 230

External sensor (4-lead technique)

Pt100

Programmer

V/mA 0 to 10 / 0 to 24

Temperature recorder (0 V = 0 °C)

mV/K 10 (RI = 100 Ohms)


Safety installations
(adjustable via LED):
High temperature protection

°C -25 ... 85

Low temperature protection

°C -25 ... 85

Low liquid level protection

float switch

Classification according to DIN 12876-1

III

Overload protection for pump motor

contactor

Overload protection for cooling compressorcontactor
Alarm signal

optical + audible


Environmental conditions according to IEC 61 010-1:

Use only indoor.

Altitude up to 2000 m - normal zero.

Ambient temperature: +5 ... +40 °C

Air

humidity:

Max. rel. humidity 80 % for temperatures up to +31 °C,

linear decrease down to 50 % relative humidity at a temperature of +40 °C

Max. mains fluctuations of ±10 % are permissible.

IP class according to IEC 60 529

IP21

The unit corresponds to Class I
Overvoltage category

II

Pollution degree

2

Caution:

The unit is not for use in explosive environment.

Standards for interference resistance according to EN 61326-1
This unit is an ISM device classified in Group 1 (using high frequency for internal purposes)
Class A (industrial and commercial range).