Appendix

p = 10

(U - 7.75) / 0.75 + c

U = 0.75 × (log p - c) + 7.75

where

U

p

c

[V]

[mbar]

0

[V]

[Pa]

2

[V]

[Torr]

-0.125

1E+01

1E–01

1E–03

1E–05

1E–07

1E–09

1E+03

Measuring signal U[V]

0.0

S

ens

or

e

rro

r

Inad

m

is

si

bl

e ran

ge

Inad

m

is

si

bl

e ran

ge

1E+04

1E+02

1E+00

1E–02

1E–04

1E–06

1E–08

1E–10

1.0

2.0

3.0

4.0

5.0

6.0

7.0

8.0

9.0

10.0

Pressure p [mbar]

A: Relationship Output

Signal – Pressure

Conversion formulae

Conversion curve