"Commands and Responses" continued:

Host Command

Host
Command syntax

BPG mode

Response syntax

RIG mode

Response syntax

Programming example and remarks

Set + threshold
(Setpoint A)

#aaSL+x.xxEsyy CR

*aa_PROGM_OK CR
(*aa_zMIN_HYS CR)

→ BPG mode

T: #02SL+1.00E–04 CR
R: *02_PROGM_OK CR

Set – threshold
(Setpoint A)

#aaSL–x.xxEsyy CR

*aa_PROGM_OK CR
(*aa_zMIN_HYS CR)

→ BPG mode

T: #02SL–2.00E–04 CR
R: *02_PROGM_OK CR (SL+

≠ SL–)

Read + threshold
(Setpoint A)

#aaRL+x.xxEsyy CR

*aa_PROGM_OK CR

→ BPG mode

T: #02SL+3.00E–04 CR
R: *02_PROGM_OK CR

Read – threshold
(Setpoint A)

#aaRL–x.xxEsyy CR

*aa_PROGM_OK CR

→ BPG mode

T: #02SL–3.00E–04 CR
R: *02_–MIN_HYS CR (SL+ = SL–)

Set + threshold
(Setpoint B)

#aaSH+x.xxEsyy CR *aa_PROGM_OK CR

(*aa_zMIN_HYS CR)

→ BPG mode

SL+/SL– and SH+/ SH– commands define the
setpoints of the switching functions and

Set – threshold–
(Setpoint B)

#aaSH–x.xxEsyy CR *aa_PROGM_OK CR

(*aa_zMIN_HYS CR)

→ BPG mode

the operation of the corresponding relays
(

→ 42).

Read + threshold
(Setpoint B)

#aaRH+x.xxEsyy CR *aa_PROGM_OK CR

→ BPG mode

Read – threshold–
(Setpoint B)

#aaRH–x.xxEsyy CR *aa_PROGM_OK CR

→ BPG mode

Read threshold
potentiometer
(Setpoint A)

#aaGT1 CR

*aa_x.xxEsyy CR

→ BPG mode

T: #02GT1 CR
R:
*02_3.50E–04 CR

Read threshold
potentiometer
(Setpoint B)

#aaGT2 CR

*aa_x.xxEsyy CR

→ BPG mode

Set factory settings
(default)

#aaFAC CR

*aa_PROGM_OK CR

→ BPG mode

All communication parameters will be set to
default values (default mode: BPG400)

Set data rate *)

#aaSBrate CR

*aa_PROGM_OK CR

→ BPG mode

T: #02SB9600 CR
R: *02_PROGM_OK CR

New data rate becomes effective after power
is cycled or a reset command is executed.

Set parity none *)

#aaSPN CR

*aa_PROGM_OK CR

→ BPG mode

New setting becomes effective after power is

Set parity odd *)

#aaSPO CR

*aa_PROGM_OK CR

→ BPG mode

cycled or a reset command is executed.

Set parity even *)

#aaSPE CR

*aa_PROGM_OK CR

→ BPG mode

Unlock

#aaUNL CR

*aa_PROGM_OK CR

→ BPG mode

T: #02UNL CR
R: *02_PROGM_OK CR

Prior to the commands SB, SPN, SPO, SPE,
SDM and GDM, a UNL command must be
executed. An attempt to execute the listed
commands without a UNL command will pro-
duce the error message
?02_COM_ERR CR

Toggle unlock

#aaTLU CR

*aa_1_UL_t CR

→ BPG mode

T: #02TLU CR
R: *02_1_UL_ON CR
T: #02TLU CR
R: *02_1_UL_OFF CR

The TLU command toggles the UNL function.
When TLU is in the ON state, the commands
SB, SPN, SPO, SPE, SDM or GDM can be
executed after a UNL command has been car-
ried out. An attempt to execute the commands
SB, SPN, SPO, SPE or GDM while TLU is in
the OFF state, an error message
?02_SYNTX_ER CR will result.

Reset

#aaRST CR

No response

No response

After a reset command, communication can be
reestablished after 3 seconds.

Set device mode *)

#aaSDMmodeT CR

*aa_PROGM_OK CR

→ BPG mode

T: #02SDM_RIG CR
R: *02_PROGM_OK CR

Get device mode *)

#aaGDM CR

*aa_modeR CR

→ BPG mode

T: #02GDM CR
R: *02_BPG_400_ CR

*) To prevent accidental changes of these parameters, a TLU - UNL sequence

has to be executed prior to the SB, SPN, SPO, SPE, SDM and GDM com-
mands (

→ "Toggle unlock" and "Toggle" commands).