INFICON BPG400 ATM to Ultra-High Vacuum Gauge User Manual
Page 48

48
tina03e1-b (2004-02) BPG400 v1.om
• Pin approx. ш1.3 Ч 50 mm (e.g. a bent open paper clip)
Gauges BPG400-SD, -SP and -SR are mechanically slightly different from the
BPG400. The adjustment opening of BPG400-SD, -SP and -SR is on one side of
the gauge housing. However, the adjustment procedure is the same for all gauge
versions.
Operate gauge for approx. 10 minutes at atmospheric pressure.
If the gauge was operated before in the BA range, a cooling-
down time of approx. 30 minutes is to be expected (gauge tem-
perature = ambient temperature).
Insert the pin through the opening and push the button inside for at least
5 s.
BPG400
BPG400-SD
BPG400-SP
BPG400-SR
max. ø1.3
Gauges with display will show the
reading "1000 mbar" and the func-
tion "A" when the button has been
pushed for 4 s. Upon completion of
the adjustment, the function indi-
cation "A" disappears.
The gauge is automatically adjusted (
≈10 s).
The gauge is now adjusted at atmospheric pressure.
A zero point adjustment is recommended
• after the sensor has been exchanged
• if display shows "FAIL 5" (→ 30)
• as part of the usual maintenance work for quality assurance
Required tools
Procedure
6.2.2 Zero Point Adjustment
- TGF10 Tracer Gas Filler (36 pages)
- Sensistor ILS500 F Leak Detection Filler (90 pages)
- T-Guard Leak Detection Sensor (85 pages)
- T-Guard Leak Detection Sensor Interface description (40 pages)
- Sensistor ISH2000P Hydrogen Leak Detector, Panel Model (51 pages)
- Sensistor ISH2000 HySpeed Hydrogen Leak Detector (54 pages)
- LDS3000 Modular Leak Detector (52 pages)
- LDS3000 Modular Leak Detector Interface description (56 pages)
- BM1000 Bus module (14 pages)
- I/O1000 I/O module (18 pages)
- CU1000 Control unit (24 pages)
- Helium Leak Detector Modul1000 (130 pages)
- Helium Leak Detector Modul1000 Interface description (40 pages)
- UL5000 Dry Helium Leak Detector (108 pages)
- UL5000 Dry Helium Leak Detector Interface description (14 pages)
- UL1000 Fab Dry Helium Leak Detector (119 pages)
- HLD6000 Refrigerant Leak Detector (76 pages)
- HLD6000 Refrigerant Leak Detector Interface Description (40 pages)
- IO1000 I/O module (18 pages)
- Ecotec E3000 Multigas-Sniffer-Leak Detector (92 pages)
- Ecotec E3000 Multigas-Sniffer-Leak Detector Interface description (36 pages)
- Sensistor XRS9012 Hydrogen Leak Detector User Manual (28 pages)
- Sensistor XRS9012 Hydrogen Leak Detector Maintenance manual (14 pages)
- Extrima Ex-certified Hydrogen Leak Detector (62 pages)
- Sensistor ILS500 Leak Detection System (107 pages)
- Sensistor ISH2000 Hydrogen Leak Detector (58 pages)
- Sensistor ISH2000 Hydrogen Leak Detector (108 pages)
- Sensistor Sentrac Hydrogen Leak Detector (86 pages)
- Protec P3000(XL) Helium Leak Detector (132 pages)
- Pilot Plus Vacuum Gauge (2 pages)
- CO Check Carbon Monoxide Meter (2 pages)
- GAS-Mate Combustible Gas Leak Detector (12 pages)
- Whisper Ultrasonic Leak Detector (8 pages)
- Vortex AC Refrigerant Recovery Machine 115V (20 pages)
- Vortex AC Refrigerant Recovery Machine 230V (16 pages)
- Wey-TEK Refrigerant Charging Scale & Optional Charging Module (44 pages)
- Wey-TEK Refrigerant Charging Scale & Optional Charging Module (2 pages)
- D-TEK CO2 Refrigerant Leak Detector (12 pages)
- TEK-Mate Refrigerant Leak Detector (12 pages)
- Compass Refrigerant Leak Detector (12 pages)
- D-TEK Select Refrigerant Leak Detector (12 pages)
- Explorer Portable Gas Chromatograph (369 pages)
- MicroFID II Portable Flame Ionization Detector (89 pages)
- DataFID Portable Flame Ionization Detector for Landfill Emissions Monitoring (91 pages)
- Hydrostik Hydrogen Fuel Cylinder Installation (7 pages)