3 syntax description – INFICON BPG400 ATM to Ultra-High Vacuum Gauge User Manual
Page 39

tina03e1-b (2004-02) BPG400 v1.om
39
In the table "Commands and Responses" (
→ 40) the following variables are
used:
Variable Description
Values, range,
aa
Operating device
address
00 … 3F
hex
fff
Emission current
_25UA
5.0MA
_20MA
= 25
µA
= 5 mA
= 20 mA
modeR
Gauge operation
mode, response
BPG_400_
RIG_
_
_
_
_
= BPG400 mode
= RIG mode (reduced ion
gauge mode)
modeT
Gauge operation
mode, command
BPG400
RIG
= BPG400 mode
= RIG mode (reduced ion
gauge mode)
n
Filament selected
1
2
No action, for future use
oo
Address offset
00 … 3F
hex
rate
Data transfer rate
300, 1200, 2400, 4800,
9600, 19'200 or 28'800 Baud
status
Gauge status
BPG mode:
BPG_ST_0
BPG_ST_5
BPG_ST_8
BPG_ST_9
RIG mode:
01_OVPRS
02_EMISS
00_ST_OK
= no error
= Pirani warning
= BA error
= Pirani error
If emission is switched off
= BA error
= normal operation
t
Toggle status
ON
OFF
= toggle function on
= toggle function off
unit
Pressure unit
MBAR, TORR, PASCAL
v.vv
Firmware version of
gauge
Three characters and a decimal point. Higher
revision numbers indicate newer versions of
firmware. Example: 1.00.
x.xx
Mantissa of
pressure values
1.00 … 9.99
syy
Signed exponent of
pressure values
s
yy
= Sign of exponent, +/
–
= Exponent
00 … 09
z
Operation of relays
(switching function)
+
–
= Relay on below threshold
= Relay on above threshold
All Commands and responses are terminated by "carriage return" (CR).
All response messages contain 13 characters, including CR.
For better readability, a space character is signified by a " _ " (underline)
character in the table below.
4.9.3 Syntax Description
4.9.3.1 Definitions, Legend