Response
Byte 0

See Subindex 0x02

Byte 1

Unused

Byte 2

0: Zeroing successful
1: Zeroing failed; out-of-range
2: Zeroing failed; cumulative out-of-range
3: Zeroing failed: measurement invalid
255: No previous Zero Adjust command issued


Execution of this command will start a Full Scale Adjust operation.

Index

SI

DataType

NV Access PM Name

FB41 0x01

V_STRING

RW Command

0x02

BYTE

RO

Status

0x03

V_STRING

RO

Response

Command
Byte 0

0: Full Scale Adjust

Byte 1

Index of the Sub Sensor Instance (1-8)

Byte 2…5 Full Scale value (Data format: REAL, always 0)

Status (supported values)

0
1
2

3

255

Last command completed, no errors, no reply available
Last command completed, no errors, reply available
Last command completed, errors present, no reply
available
Last command completed, errors present, reply available
Command is executing

Response
Byte 0

See Subindex 0x02

Byte 1

Unused

Byte 2

0: Full Scale Adjust successful
1: Full Scale Adjust failed: out-of-range
255: No previous Full Scale Adjust command issued


Execution of this command causes the device to emulate a complete power cycle.
This includes an ESC reset. An SDP may limit some behavior of the power cycle
emulation, but shall not exclude the EtherCAT interface.

As consequence of an ESC reset all following devices are disconnected

from the network.

There are two versions of this command:
• Standard reset (as described above)
• Factory reset (as described above, but additionally, all parameters are restored

to as-shipped defaults).

Index

SI

DataType

NV Access PM Name

FBF0 0x01

V_STRING

RW Command

0x02

BYTE

RO

Status

0x03

V_STRING

RO

Response

Subindex 0x03

4.18.2

Full Scale Adjust

Subindex 0x01

Subindex 0x02

Subindex 0x03

4.18.3 Device

Reset

Command