INFICON PSG55x ATM to Medium Vacuum Gauge (EtherCAT) User Manual
Page 16

16
tira85e1 (2013-06) PxG55x.cp
• PCG gauges with or without ATM: The "Manufacturer Configuration Module 1"
is assigned to the CDG module.
• PSG gauges: The "Manufacturer Configuration Module 1" is assigned to the
Pirani module.
Index
SI
DataType
NV Access PM Name
2800 0x01
BYTE RW Safe
State
0x02
REAL
RW
Safe
Value
Specifies the behavior for the value for States other than Valid.
Safe State
0
1
2
3
Zero
Full Scale
Hold last value
Use safe value
Safe Value: The value to be used for Safe State = Safe Value.
• PCG gauges with ATM: The "Manufacturer Configuration Module 2" is assigned
to the ATM module.
• PCG gauges without ATM: The "Manufacturer Configuration Module 2" is
assigned to the Pirani module.
Index
SI
DataType
NV Access PM Name
2801 0x01
BYTE RW Safe
State
0x02
REAL
RW
Safe
Value
Specifies the behavior for the Value for states other than Valid.
Safe State
0
1
2
3
Zero
Full Scale
Hold last value
Use safe value
Safe Value: The value to be used for Safe State = Safe Value.
• PCG gauges with ATM: The "Manufacturer Configuration Module 3" is assigned
to the Pirani module.
Index
SI
DataType
NV Access PM Name
2802 0x01
BYTE RW Safe
State
0x02
REAL
RW
Safe
Value
4.3.2 Manufacturer
Configuration Module 1
Subindex 0x01
Subindex 0x02
4.3.3 Manufacturer
Configuration Module 2
(PCG with or without
ATM only)
Subindex 0x01
Subindex 0x02
4.3.4 Manufacturer
Configuration Module 3
(PCG with ATM only)