beautypg.com

3 input common, 4 input piezo, 5 input common – INFICON PSG55x ATM to Medium Vacuum Gauge (EtherCAT) User Manual

Page 18

background image

18

tira85e1 (2013-06) PxG55x.cp

Indicates whether the Value parameter contains a value in under range.

Reading Valid

0

No Underrange Exceeded

1 Underrange

Exceeded


Index

SI

DataType

NV Access PM Name

6010 0x0E

BOOL RO tx

TxPdoState

0x11

REAL

RO

tx

Sensor

Value

Is set if the device is not in Safe State (value (I 0x6nn0, SI 0x11) = valid)

TxPdoState

0
1

Invalid
Valid

The corrected, converted, calibrated final analog input value of the sensor.

Index

SI

DataType

NV Access PM Name

6012 0x01

BOOL RO tx

Reading

Valid

0x02

BOOL

RO

tx

Overrange

Exceeded

0x03

BOOL

RO

tx

Underrange

Exceeded

Indicates whether the Value parameter contains a valid value within the specified
accuracy or not.

Reading Valid

0 Invalid
1 Valid


Indicates whether the Value parameter contains a value in over range.

Reading Valid

0

No Overrange Exceeded

1 Overrange

Exceeded


Indicates whether the Value parameter contains a value in under range.

Reading Valid

0

No Underrange Exceeded

1 Underrange

Exceeded


Index

SI

DataType

NV Access PM Name

6020 0x0E

BOOL RO tx

TxPdoState

0x11

REAL

RO

tx

Sensor

Value

Subindex 0x03

4.4.3 Input Common

Subindex 0x0E

Subindex 0x11

4.4.4 Input Piezo

Subindex 0x01

Subindex 0x02

Subindex 0x03

4.4.5 Input Common