beautypg.com

INFICON PSG55x ATM to Medium Vacuum Gauge (EtherCAT) User Manual

Page 35

background image

tira85e1 (2013-06) PxG55x.cp

35

Index

SI

DataType

NV Access PM Name

9001 0x01

UINT

RO Sensor

Warning

0x02

UINT

RO

Sensor

Alarm

Sensor Warnings
Bit 0
Bit 1
Bit 2…15

0
Electronics Warning
0

Sensor Alarms
Bit 0
Bit 1
Bit 3…15

Diaphragm Failure
Electronics Failure
0


Index

SI

DataType

NV Access PM Name

9010 0x02

REAL x

RO Highest

Informational

Measurement Value

0x03

REAL

x

RO

Highest

Precision

Measurement Value

0x04

REAL

x

RO

Lowest

Precision

Measurement Value

Highest Informational Measurement Value: Highest value that the gauge can
measure without a specified accuracy.

Highest Precision Measurement Value: Highest valid value at which the gauge is
specified with an accuracy value. Above that value, the parameter Reading Valid is
set to invalid.

Lowest Precision Measurement Value: Lowest valid value at which the gauge is
specified with an accuracy value. Below that value, the parameter Reading Valid is
set to invalid.

Index

SI

DataType

NV Access PM Name

9012 0x01

UINT

RO Sensor

Warning

0x02

UINT

RO

Sensor

Alarm

Sensor Warnings
Bit 0…15

0

Sensor Alarms
Bit 0…15

0


Index

SI

DataType

NV Access PM Name

9020 0x02

REAL x

RO Highest

Informational

Measurement Value

0x03

REAL

x

RO

Highest

Precision

Measurement Value

0x04

REAL

x

RO

Lowest

Precision

Measurement Value

4.10.1 Information

Capacitance Diaphragm

Subindex 0x01

Subindex 0x02

4.10.2 Information

Common

Subindex 0x02

Subindex 0x03

Subindex 0x04

4.10.3 Information

Piezo

Subindex 0x01

Subindex 0x02

4.10.4 Information

Common