INFICON PSG55x ATM to Medium Vacuum Gauge (EtherCAT) User Manual
Page 35

tira85e1 (2013-06) PxG55x.cp
35
Index
SI
DataType
NV Access PM Name
9001 0x01
UINT
RO Sensor
Warning
0x02
UINT
RO
Sensor
Alarm
Sensor Warnings
Bit 0
Bit 1
Bit 2…15
0
Electronics Warning
0
Sensor Alarms
Bit 0
Bit 1
Bit 3…15
Diaphragm Failure
Electronics Failure
0
Index
SI
DataType
NV Access PM Name
9010 0x02
REAL x
RO Highest
Informational
Measurement Value
0x03
REAL
x
RO
Highest
Precision
Measurement Value
0x04
REAL
x
RO
Lowest
Precision
Measurement Value
Highest Informational Measurement Value: Highest value that the gauge can
measure without a specified accuracy.
Highest Precision Measurement Value: Highest valid value at which the gauge is
specified with an accuracy value. Above that value, the parameter Reading Valid is
set to invalid.
Lowest Precision Measurement Value: Lowest valid value at which the gauge is
specified with an accuracy value. Below that value, the parameter Reading Valid is
set to invalid.
Index
SI
DataType
NV Access PM Name
9012 0x01
UINT
RO Sensor
Warning
0x02
UINT
RO
Sensor
Alarm
Sensor Warnings
Bit 0…15
0
Sensor Alarms
Bit 0…15
0
Index
SI
DataType
NV Access PM Name
9020 0x02
REAL x
RO Highest
Informational
Measurement Value
0x03
REAL
x
RO
Highest
Precision
Measurement Value
0x04
REAL
x
RO
Lowest
Precision
Measurement Value
4.10.1 Information
Capacitance Diaphragm
Subindex 0x01
Subindex 0x02
4.10.2 Information
Common
Subindex 0x02
Subindex 0x03
Subindex 0x04
4.10.3 Information
Piezo
Subindex 0x01
Subindex 0x02
4.10.4 Information
Common