High Trip Point limit: High limit to trigger trip point condition if Input Value (I 0x900F,
SI 0x01) is above this limit.

Object index of High Trip Point Value source.

Bit 16…31: Index
Bit 08…15: Subindex
Bit 00…07: reserved

For a fix limit set value to 0x800F1100 (High Trip Point Limit).

For ATM Mode set value to 0x800F1300 (Percentage High Trip Source).

Percentage of High Trip Source: Percentage of Piezo Sensor Value (I 0x60010, SI
0x11) in [%].

Low Trip Point Limit: Low limit to trigger trip point condition if Input Value (I 0x900F,
SI 0x01) is below this limit.

Object index of Low Trip Point Value source.

Bit 16…31: Index
Bit 08…15: Subindex
Bit 00…07: reserved

For a fix limit set value to 0x800F1400.

For ATM Mode set value to 0x800F1600.

High Trip Point Limit: High limit to trigger trip point condition if Input Value (I
0x900F, SI 0x01) is above this limit.

High Trip Hysteresis: Hysteresis value for High Trip Point.

Low Trip Hysteresis: Hysteresis value for Low Trip Point.

Object index of active source of (I 0x900F SI 0x01) Input Value.

Source Index
Bit 16…31
Bit 08…15
Bit 00…07

Index
Subindex
Reserved


This chapter describes the configuration area for the PSG.

The High Trip Point is calculated by:

High Trip Point = Value

*)

× Percentage

**)

*)

Value from Piezo Sensor (I: 0x6010, SI: 0x11)

**)

Percentage

High Trip Source (I: 0x800E/F, SI: 0x13)

In this case, the High Trip Point value is automatically updated by the device itself.
Any user-written value to the parameter High Trip will be refused with an abort
code.

The Low Trip Point is a fixed value:

Subindex 0x11

Subindex 0x12

Subindex 0x13

Subindex 0x14

Subindex 0x15

Subindex 0x16

Subindex 0x17

Subindex 0x18

Subindex 0x1A

4.9 PSG Configuration Area

(0x8000…0x8FFF)

4.9.1 Configuration Trip

Point 1