tira58e1-a (2012-03) PxG5xx.cp

Following is the instance 3 with the subclass extension of the Atmosphere Pressure
Sensor of the gauge (valid for PCG55x).

Attribute

ID

Access rule NV/

V

Name

DeviceNet
data type

Description of attribute

Semantics of values

3

"Semantics"

NV Data type USINT

Determines the data type
of value and all related
attributes as specified in
this table.

→ "Semantics" section below
[default] = INT

4

"Semantics"

NV Data units UINT

Determines the units con-
text of value and all related
attributes.

[default] = Counts

5 Get

V

Reading
valid

BOOL

Indicates that the value
attribute contains a valid
value.

0 = invalid
1 = valid
(invalid: e.g., sensor is out of its
specified range)

6

Get

V Value

INT or REAL

Analog input value.

The corrected, converted, calibrated
final value of the sensor.
→ "Semantics" section

7

Get

V Status

BYTE

Alarm and warning state of
this object instance.

= 0

10

Get

NV Full scale

INT or REAL

The value of full scale for
the sensor.

[default] = maximum allowable value
for the data type

25

Set

NV Safe state USINT

Specifies the behavior for
the value for states other
than execute.

→ "Semantics" section below
[default] = 0

26

Set

NV Safe value INT or speci-

fied by data
type

The Value to be used for
safe state = safe value

→ "Semantics" section below
[default] = 0

32

Get

NV Overrange INT or REAL

Specifies the highest valid
value.

The value above which attribute
reading valid

is set to invalid.

[default] = maximum allowable value
for the data type

33 Get

NV

Under-
range

INT or REAL

Specifies the lowest valid
value.

The value below which attribute
reading valid

is set to invalid.

[default] = minimum allowable value
for the data type

94 Get

V

Sensor
warning

Structure of
byte

Bit definitions of sensor
warnings.

default = 0
→ "Semantics" section below

95 Get

V

Sensor
alarm

Structure of
byte

Bit definitions of sensor
alarms.

default = 0
→ "Semantics" section below

96 Get

V

Status
extension

Byte

Bit-mapped byte providing
additional status bits of an
S-Analog sensor instance.

default = 0
→ "Semantics" section below

99

Get

V Subclass

UINT

Identifies a subset of addi-
tional instance attributes,
services and behaviors.

3 = Capacitance Manometer


4.4.3.10 PCG55x - Instance