INFICON PSG55x ATM to Medium Vacuum Gauge (DeviceNet) User Manual
Page 30
![background image](/manuals/562585/30/background.png)
30
tira58e1-a (2012-03) PxG5xx.cp
Following is the instance 3 with the subclass extension of the Atmosphere Pressure
Sensor of the gauge (valid for PCG55x).
Attribute
ID
Access rule NV/
V
Name
DeviceNet
data type
Description of attribute
Semantics of values
3
→
"Semantics"
NV Data type USINT
Determines the data type
of value and all related
attributes as specified in
this table.
→ "Semantics" section below
[default] = INT
4
→
"Semantics"
NV Data units UINT
Determines the units con-
text of value and all related
attributes.
[default] = Counts
5 Get
V
Reading
valid
BOOL
Indicates that the value
attribute contains a valid
value.
0 = invalid
1 = valid
(invalid: e.g., sensor is out of its
specified range)
6
Get
V Value
INT or REAL
Analog input value.
The corrected, converted, calibrated
final value of the sensor.
→ "Semantics" section
7
Get
V Status
BYTE
Alarm and warning state of
this object instance.
= 0
10
Get
NV Full scale
INT or REAL
The value of full scale for
the sensor.
[default] = maximum allowable value
for the data type
25
Set
NV Safe state USINT
Specifies the behavior for
the value for states other
than execute.
→ "Semantics" section below
[default] = 0
26
Set
NV Safe value INT or speci-
fied by data
type
The Value to be used for
safe state = safe value
→ "Semantics" section below
[default] = 0
32
Get
NV Overrange INT or REAL
Specifies the highest valid
value.
The value above which attribute
reading valid
is set to invalid.
[default] = maximum allowable value
for the data type
33 Get
NV
Under-
range
INT or REAL
Specifies the lowest valid
value.
The value below which attribute
reading valid
is set to invalid.
[default] = minimum allowable value
for the data type
94 Get
V
Sensor
warning
Structure of
byte
Bit definitions of sensor
warnings.
default = 0
→ "Semantics" section below
95 Get
V
Sensor
alarm
Structure of
byte
Bit definitions of sensor
alarms.
default = 0
→ "Semantics" section below
96 Get
V
Status
extension
Byte
Bit-mapped byte providing
additional status bits of an
S-Analog sensor instance.
default = 0
→ "Semantics" section below
99
Get
V Subclass
UINT
Identifies a subset of addi-
tional instance attributes,
services and behaviors.
3 = Capacitance Manometer
4.4.3.10 PCG55x - Instance
Attributes of
Instance 3