INFICON PSG55x ATM to Medium Vacuum Gauge (DeviceNet) User Manual
Page 15
tira58e1-a (2012-03) PxG5xx.cp
15
Attribute ID
Access rule
NV/
V
Name
DeviceNet data type
Description of attribute
14 Get
V
Exception detail
warning
STRUCT of:
A structure of three structures containing
a bit mapped representation of the
warning detail
Common exception
detail
STRUCT of:
Size 2
USINT
Number of common detail bytes
Detail 0
ARRAY of:
→ "Semantics" section below.
Detail
1 BYTE
→ "Semantics" section below.
Device exception
detail
STRUCT of:
Size
USINT
Number of device detail bytes
→ "Semantics" section below.
Detail
0 BYTE
→ "Semantics" section below.
Detail
1 BYTE
→ "Semantics" section below.
Detail
2 BYTE
→ "Semantics" section below.
Detail
3 BYTE
→ "Semantics" section below.
Detail
4 BYTE
→ "Semantics" section below.
Manufacturer
exception detail
STRUCT of:
Size 1
USINT
Number of manufacturer detail bytes = 1
Detail 0
ARRAY of:
Manufacturer specified
15 Set
NV
Alarm
enable
BOOL
→ "Semantics" section below.
16 Set
NV
Warning
enable
BOOL
→ "Semantics" section below.
101 Get
NV
Software
revision
base unit
SHORT STRING
ASCII text, for example "010100"
102
Get
V
Run hours
UINT
1 = 1 hour