7 semantics of s-analog sensor instance 2 – INFICON PSG55x ATM to Medium Vacuum Gauge (DeviceNet) User Manual

Page 28

background image

28

tira58e1-a (2012-03) PxG5xx.cp



→ Instance 1

→ Instance 1

→ Instance 1

→ Instance 1

→ Instance 1

16 bits are used as sensor warnings.


Data component

Bit 7

Bit 6

Bit 5

Bit 4

Bit 3

Bit 2

Bit 1

Bit 0

Sensor warning byte 0

0 0 0 0 0 0 0 0

Sensor warning byte 1

0 0 0 0 0 0

Electronics

Warning

0

16 Bits are used as sensor faults.

Data component

Bit 7

Bit 6

Bit 5

Bit 4

Bit 3

Bit 2

Bit 1

Bit 0

Sensor alarm byte 0

0 0 0 0 0 0 0

Sensor

Element

Failure

Sensor alarm byte 1

0 0 0 0 0 0

Electronics

Failure

0


8 bits providing the current sensor alarm state of the instance

Data component

Bit 7

Bit 6

Bit 5

Bit 4

Bit 3

Bit 2

Bit 1

Bit 0

Sensor alarm byte 0

0 0 0 0 0

Underrange

exceeded

Overrange

exceeded

Reading
invalid

1)

1)

Logical inversion of reading valid


The S-Analog sensor object provides the following common services:

Service

code

Service name

Description of service

0E

h

Get_Attribute_Single

Returns the contents of the specified attribute.

10

h

Set_Attribute_Single

Modifies an attribute value.


Service

Code

Service Name

Description of Service

4C

h

Gain Adjust

Performs an Atmosphere Adjust for the Pirani-
Sensor

4B

h

Zero Adjust

Performs an High Vaccum Adjust for the
Pirani-Sensor

Parameter

Data Type

Description Semantics of Values

Gain Adjust parameter UINT or

FLOAT
depending
on data type

The target
value for the
zero
calibration

The value to which the
attribute 6 (value)will be set

4.4.3.7 Semantics of S-Analog

Sensor Instance 2

Data type

Data units

Value

Safe state

Safe value

Sensor warning

Sensor alarm

Status extension

4.4.3.8 Common Services of

S-Analog Sensor
Instance 2

4.4.3.9 User Adjust Service

Gain Adjust Request
Service Data Field
Parameters