2 identity object – INFICON PSG55x ATM to Medium Vacuum Gauge (DeviceNet) User Manual
Page 12
12
tira58e1-a (2012-03) PxG5xx.cp
Class Code 01
h
= 01
d
This object provides identification of and general information about the device.
Attribute ID Access rule
Name
DeviceNet
data type
Description of attribute
Semantics of values
1
Get
Revision
UINT
Revision of this object.
The current value assigned to this
attribute is one (01). If updates that
require an increase in this value are
made, then the value of this attribute
increases by one.
2
Get
Max instance UINT
Maximum instance number of
an object currently created in
this class level of the device.
The largest Instance number of a
created object at this class hierarchy
level.
Attribute ID Access rule
NV/V Name
DeviceNet data type
Description of attribute
1
Get
NV
Vendor ID
UINT
Manufacturer identification
Value: 633 (INFICON AG) (default)
2 Get
NV
Device
type
UINT
1C
h
= Vacuum pressure gauge
3 Get
NV
Product
code
UINT
→ "Semantics" section below
4
Get
NV
Revision
STRUCT
Major Revision 1; Minor Revision 1
5 Get
NV
Status
WORD
6 Get
NV
Serial
number
UDINT
7
Get
NV
Product name
SHORT STRING
→ "Semantics" section below
100 Get
NV
Software
Revision
Level
SHORT STRING
Revision of the DeviceNet board software
(major number) (minor number)
(development number)
e.g. 01 02 00
Relationship between product code and product name.
Product code Sensor type
Product name
19
CDG, Pirani
PCG550, PCG552, PCG554
20
CDG, Pirani, ATM PCG550, PCG552, PCG554
21
Pirani
PSG550, PSG552, PSG554
Service Code Name
Description
5 (05
h
)
Reset
0 = last installation,
1 = default installation (factory reset)
14 (0E
h
) Get_Attribute_Single
4.2 Identity Object
4.2.1 Class Attributes
4.2.2 Instance Attributes
4.2.3 Semantics
4.2.4 Services