5.4.16 Command: W

Parameters: . . . . . . . . . . . . . . . . . . . None

Description: . . . . . . . . . . . . . . . . . . . Returns the rate, thickness & frequency for

each sensor simultaneously.

Example: . . . . . . . . . . . . . . . . . . . . . W 0.0 00.78 58.654 5497324.373 00.78

58.668 5497433.978 00.78 58.648
5500330.091 00.79 58.654 5500440.187
00.77 58.655 5500550.287 00.78 58.662
5500660.020

In order: ignore the first 0.00 in the

response. Next: Sensor 1 Rate (Å/s), Sensor 1
Thickness (Å), Sensor 1 Freqency (Hz), Sensor 2 Rate
(Å/s) Sensor 2 Thickness (Å), Sensor 2 Frequency
(Hz), etc.

5.4.17 Command: Y

Parameters: . . . . . . . . . . . . . . . . . . . None.

Description: . . . . . . . . . . . . . . . . . . . Read the Power-Up Reset flag. The

Power-Up Reset flag is set during boot-up of
the SQM-160 and stays set until read
through the RS-232 interface. After the flag is
read, it is reset and will not be set again until
the SQM-160 is power cycled.

Example: . . . . . . . . . . . . . . . . . . . . . Y A1

Power-Up Reset flag is set.

Y A0

Power-Up Reset flag is reset.

5.4.18 Command: Z

Parameters: . . . . . . . . . . . . . . . . . . . None.

Description: . . . . . . . . . . . . . . . . . . . Set all Film and System menu parameters to

defaults. Note that this command can take
over 1 second to complete

Example: . . . . . . . . . . . . . . . . . . . . . Z A

All Film and System parameters are set to

defaults.