3.18 Relay Operation

The four relays of the SQM-160 are physically single-pole, normally-open
(1FormA) relays. However, each can be programmed to act as either
normally-open or normally-closed during SQM-160 operation. It is important to
keep in mind that all relays will open if the SQM-160 is turned off or loses power.
See

section 3.20 on page 3-16

for relay wiring.

To set the relay operating mode:

1

Press Program to enter Program mode.

2

Press Prev to enter the System Menu.

3

Press Next until RELAYS is shown and then press the Control Knob to display
Relay 1.

4

Turn the Control Knob left or right to select NO (normally open) or NC (normally
closed). Press the Control Knob to accept the value and display the next relay.

5

Repeat Step 4 for each of the four relays.

6

Press Program to exit the System Menu and return to normal mode.