10. Messages

10.1. ERROR MESSAGES
If an error is detected, a corresponding error message appears that must be
acknowledged by pressing any key. Possible error messages are as follows:

E.01

The switching points and hysteresis that have been set are such that the
resulting switch-back point falls outside the permissible setting range.
Example:
Switching point has been set to 8.20 inch, the hysteresis to 9.05 inch.

Action:

Correct the setting(s).

E.10

A data error has been detected in the saved settings. Possible causes
are strong electromagnetic interference or a defective component.

Action:

Check all settings (programming enable, switching points, switch-back

points and basic settings) and correct these if necessary. If this error
occurs frequently, please contact our service department.

E.20

A communication error occurring within the unit has been detected.
Possible causes are strong electromagnetic interference or a
defective component.

Action:

Acknowledge the error message by pressing any key. If the error
message continues to be displayed, please contact our service
department.