HYDAC ENS 3000 User Manual
Page 15

E
15
Setting
Display Settings
available
Default
setting
Primary display (Primary)
Value to be continuously
displayed:
niv
Actual fluid level
Peak fluid level
Lowest fluid level
Actual temperature
Peak temperature
Lowest temperature
to
4 Switching point 1 to 4
Display off
niv/
n.Top/
n.Min/
Temp/
T.Top/
T.Min/
S.P.1/
S.P.2/
S.P.3/
S.P.4/
OFF
(See Point 5. "Digital Display")
Setting the unit of measurement for fluid level
(NIVEAU RANGE i.e. level range)
Fluid level will be displayed
in inches
inch/
cm
inch
(depending
on version)
Fluid level will be displayed
in cm
Setting the level offset (NIVEAU OFFSET) while the fluid level is displayed
The distance between the base of the tank
and the start of the active zone of the fluid
level sensor is regarded as the "offset".
Inputting this distance will generate a
reference to actual fluid levels, enabling the
actual fluid level to be displayed.
0
Setting the temperature unit (TEMPERATURE RANGE)
Temperature will be
displayed in °F
Fahr/ celc
Temperature will be
displayed in °C
Fahr
(depending
on version)