Technical data, Appendix – Heidolph Hei-VAP Precision User Manual
Page 55
Appendix
108
109
Appendix
Technical data
Base unit
Hei-VAP
Value,
Hei-VAP
Value
Digital
Base unit
Hei-VAP
HL
Base unit
Hei-VAP
ML
Base unit
Hei-VAP
HL
Base unit
Hei-VAP
ML
Dimensions (WxDxH) of drive unit in
the lowest position without a glassware
set ****
393 x 449
x 427 mm
393 x 449
x 427 mm
393 x 411
x 427 mm
393 x 449
x 427 mm
393 x 411
x 427 mm
Connected load
1400 W
Connected voltage
230 V / 50/60 Hz oder 115 V / 50/60 Hz
Drive unit
EC motor with electronic speed control
Drive unit of lift
Manual
Manual
Motor
Manual
Motor
Speed range
10 – 280 min-1
Weight incl. heating bath
without glassware set
16 kg
16 kg
16 kg
17 kg
17 kg
Lifting speed
manual
manual
approx. 40
mm/s
manual
approx. 40
mm/s
Length of stroke
155 mm
Cooling surface
1200 cm²
Panel
Value,
Value
Digital
Panel
Advantage-
HL
Panel
Advantage-
ML
Panel
Precision-
HL
Panel
Precision-
ML
Connected load
24 V dc
Connected voltage
1 W
2 W
Display
Scale
3,5” LCD
Display
3,5” LCD
Display
4,3” LCD
Farbdisplay
4,3” LCD
Farbdisplay
Display
heating bath temp. *****
Scale /
digital
3,5” LCD
screen
3,5” LCD
screen
4,3” LCD
color screen
4,3” LCD
color screen
Control panel
‒
Lift
‒
‒
yes
‒
yes
‒
Start rotation
yes
yes
yes
yes
yes
‒
Start vacuum
‒
‒
‒
yes
yes
‒
Start heating bath
yes
yes
yes
yes
yes
‒
Start timer
‒
yes
yes
yes
yes
Heating Bath Hei-VAP
Heating capacity
1300 W
Diameter of heating bath
255 mm
Heating bath material
V4A (1.4404)
Heating bath temperature range
20 – 210 °C
Heating bath control accuracy
±1 K
Heating bath temperature control
electronic/digital
Degree of protection
IP 20
Airborne sound level
Clearly lower than 85 db(A)
Permissible ambient conditions
5 – 31 °C at 80 % relative humidity 32 – 40 °C decreasing linear to a
max. rel. humidity of 50%
0 - 2000m absolute altitude
Pollution Degree 2
Installation Category II
Connected voltage *** ±10%
230 V /
50/60 Hz
230 V /
50/60 Hz
230 V /
50/60 Hz
230 V /
50/60 Hz
230 V /
50/60 Hz
Evaporation rate (L/h) ΔT* 20/40 °C
115 V /
50/60 Hz
115 V /
50/60 Hz
115 V /
50/60 Hz
115 V /
50/60 Hz
115 V /
50/60 Hz
‒
Toluene
4,1 / 8,2
4,1 / 8,2
4,1 / 8,2
4,1 / 8,2
4,1 / 8,2
‒
Acetone
2,4 / 4,7
2,4 / 4,7
2,4 / 4,7
2,4 / 4,7
2,4 / 4,7
‒
Ethanol
1,6 / 3,2
1,6 / 3,2
1,6 / 3,2
1,6 / 3,2
1,6 / 3,2
‒
Water
0,5 / 1,0
0,5 / 1,0
0,5 / 1,0
0,5 / 1,0
0,5 / 1,0
* ΔT = Difference between the heating bath temperature and the vapor temperature
** Only possible in conjunction with vacuum systems
*** Standard 230 V/50/60 Hz: other connecting voltages upon request
**** Ready to connect including the power plug
***** Hei-VAP Value Digital: digital display
Table 11-1: Technical data