IKA C 5000 control Package 2/12 User Manual
Page 41

IKA
-WERKE C 5000 control/duo-control Ver. 10 04.07
3DJH
&RQQHFWLRQ R[\JHQ VXSSO\
Installation
The 2 m long O
2
hose supplied with the system is designed for a maximum pressure
of 40 bar at room temperature. This can either be fed directly to the oxygen cylinder
or to another oxygen connection point with a pressure reducer (smallest permitted
bending radius 80 mm).
The C 29 reducer valve (accessory) for the oxy-
gen cylinder features an R ¼ inch thread for con-
necting the pressure hose.
The pressure hose is supplied with an adapter for
American pressure reducers.
i
Make sure that the oxygen supply line is connected to the C 5000 calorimeter as
described under Point
e
. Then connect the O
2
line to the laboratory oxygen supply
end.
7KH SUHVVXUH RI WKH R[\JHQ VKRXOG EH EDU EXW PXVW QRW LQ DQ\ FDVH H[FHHG
EDU Keep all gas lines and screw connections that carry oxygen free from grease. Observe the accident prevention requirements applicable to the activity and the work station. Wear personal safety equipment. Close the main valve on the oxygen supply when work is complete. Only carry out maintenance work when the system is depressurised. ☞