IKA C 5000 control Package 2/12 User Manual
Page 103

IKA
-WERKE C 5000 control/duo-control Ver. 10 04.07
3DJH
•
1R LQFUHDVH LQ WHPSHUDWXUH DIWHU HOHFWULFDO LJQLWLRQ
0HVVDJHV RQ WKH GLVSOD\
1R LQFUHDVH LQ WHPSHUDWXUH
The system displays the malfunction and interrupts the measurement. Check the
ignition wire, the cotton thread, the fuel sample and the O
2
supply. It may be that
you will have to use a combustion aid.
•
7KH FRPEXVWLRQ H[SHULPHQW ODVWV WRR ORQJ
0HVVDJHV RQ WKH GLVSOD\
3UHOLPLQDU\ H[SHULPHQW ! PLQXWHV
0DLQ H[SHULPHQW ! PLQXWHV
Adiabatic and isoperibolic measurements are interrupted after 13 minutes in the
preliminary experiment (warm-up) and 16 minutes in the main experiment. Check
the stirrer drive (rotating stirring rod after the inner vessel has been emptied), the
seal of the decomposition vessel (see C 5010 / C 5012 Operating Instructions),
the function of the cooling unit as indicated by heat escaping from the cooler
fans.
•
(UURU ZKLOH UHFRUGLQJ WHPSHUDWXUH
0HVVDJHV RQ WKH GLVSOD\
(UURU WHPS PHDV
The malfunction is displayed in the header line of the display. An acoustic signal
is heard at the same time. The temperature display stops changing. You can try
to eliminate the error with the Temp-init. command in the
0DLQWHQDQFH
menu. If
this does not work, the calorimeter must be restarted.
•
:DWHU VHQVRU HUURU
0HVVDJHV RQ WKH GLVSOD\
:DWHU VHQVRU HUURU
Please contact your Technical Service Department.
•
3UREOHP ZLWK LJQLWLRQ ZLUH
0HVVDJHV RQ WKH GLVSOD\
3UREOHP ZLWK LJQLWLRQ ZLUH
This error message is displayed if the ignition capability can no longer be ensured
during the measurement. The experiment is interrupted. Check the ignition wire,
the ignition wire fastening and the contact spring on the filling head of the inner
cover.
•
6DPSOH UDFN LQYDOLG
0HVVDJHV RQ WKH GLVSOD\
6DPSOH UDFN LQYDOLG
For more information on this message, see the Operating Instructions for the
C 5020 sample rack.