Make sure that the oxygen supply line is connected to the C 5000 calorimeter as
described under Point
d
. Then connect the O
2
line to the laboratory oxygen supply
end.
Please refer to chapter 8.4 for details how to connect to oxygen supplies.
The oxygen supply line supplied is suitable for a pressure of max. 40 bar. It is 2 m
long, and the smallest permissible bending radius is 80 mm. The pressure-reducing
valve C 29 (accessory) is suitable for connecting the O
2
line to the oxygen supply. It
has a 1/4" BSP thread for connecting the hose. For use with an American pressure-
reducer with a 1/4" NPT thread, a suitable adapter for the oxygen line is included.
g
Place the C 5001 cooling system next to the measurement cell. Push the cooling
system all the way up to the measurement cell. The joining piece on the C 5001
cooling system fits into the opening on the measurement cell that lines up with it.
Fasten the two components in place by screwing in the two counter-sunk socket
bolts. Insert the hoses into the water connections of the measurement cell.