Page 101

IKA
-WERKE C 5000 control/duo-control Ver. 10 04.07
3DJH
7URXEOHVKRRWLQJ
The C 5000 calorimeter system is subject to strict quality control during manufac-
turing. If improper functionality should nevertheless occur, you will find a series of
malfunction situations with the appropriate measures for eliminating the problem.
Most malfunctions are displayed in the header line of the display. Alternatively or in
addition, a message box may appear which the user must acknowledge. If your at-
tempts at eliminating problems are unsuccessful, please contact your authorized
IKA
Technical Service Department.
0DLQWHQDQFH PHQX
¢¡¤£
¥§¦
¨©¥¡©¥¤¨¨¤¥
The maintenance menu offers the possibility of performing a series of system func-
tions in the case of a malfunction. The commands of the maintenance menu can
only be performed if the measurement cell is in maintenance status.
The functions
2SHQ 0& &ORVH 0& )LOO ,9 (PSW\ ,9 2 )LOO 'HSUHVVXULVH
are acti-
vated by the corresponding menu command and end automatically. While the func-
tion is being performed, the menu item in question is locked.
2SHQ 0&
Opens the measurement cell cover.
&ORVH 0&
Closes the measurement cell cover.
,QIR
An information window is shown /hidden. The function can also be
called up or closed with the “.” key. You can print the information
window with the Space bar .
)LOO ,9
The inner vessel is filled with water. The filling process ends
automatically after about 120 seconds.
(PSW\ ,9
The water is pumped out of the inner vessel. The emptying ends
automatically when the inner vessel is empty.
7HPS,QLW
The temperature measurement is reinitialized.
5HVHW
Reset of the temperature control of the outer vessel.