IKA C 5000 control Package 2/12 User Manual

Page 104

background image

7URXEOHVKRRWLQJ

IKA

-WERKE C 5000 control/duo-control Ver. 10 04.07

3DJH

7HPSHUDWXUH FRQWURO RI WKH V\VWHP XQVWDEOH

0HVVDJHV RQ WKH GLVSOD\

³8QVWDEOH´ VWDWXV

The “unstable” condition may last up to 10 minutes after turning on the machine,

and up to 5 minutes between measurements. If this time is exceeded signifi-

cantly, or if the “stable” condition is no longer reached at all, select the

5HVHW

op-

tion from the

PDLQWHQDQFH

menu.

Check in addition the functionality of the cooler (make sure hot air is coming out

of the outlet). If the “Stable” status is not reached within an additional 30 minutes,

turn the machine off. If the problem persists after the unit has been turned back

on again, please contact your Technical Service Department.

0HPRU\ LV WRR ORZ

0HVVDJHV RQ WKH GLVSOD\

0HPRU\ LV WRR ORZ

The total number of all measurements that the C 5000 can manage during a run

sequence is limited to about 240. This limitation is reached as early as after 50

hours of continuous operation (25 hours with the duo-control), and the message

above appears. Afterwards, a defined exit must be carried out (menu item

(;,7

).

When you switch on again, normal operation can be continued.

This number has nothing to do with the storage capacity of the library.

$ PDOIXQFWLRQ VLWXDWLRQ ZLWKRXW D GLUHFW PHVVDJH RQ WKH GLVSOD\

/RVV RI SRZHU FRQWUROOHU LQ DQ XQGHILQHG VWDWH

If no measurement was running, the system can be restarted by turning it off and

back on again. If the system was in the process of performing a measurement,

proceed as follows:

1. Turn the system off and back on, and start normally.

2. If there is residue in the inner vessel, confirm

(PSW\ ,9

3. Open the

0DLQWHQDQFH

menu

4. Bleed the over-pressure from the decomposition vessel with

'HSUHVVXULVH

2

ILOOLQJ GRHV QRW ZRUN

Check the O

2

supply to the unit (30 bar). If there is a detectable loss of seal dur-

ing the filling procedure, replace the O

2

seal (see Section 13, “Care and mainte-

nance”).

2

YHQWLODWLRQ GRHV QRW ZRUN

Check the O

2

supply to the unit (30 bar). Check the way the ventilation hose is

installed and laid out. There must be no kinks or obstructions in the degassing

hose. Check the settings in the

&RQILJ 6HWWLQJV

menu. The

'HFRPSRVLWLRQ

item

must not be set, since this turns off O

2

ventilation.