&DORULPHWULF PHDVXUHPHQWV
IKA
-WERKE C 5000 control/duo-control Ver. 10 04.07
3DJH
Most liquid substances can be weighed directly into the crucible. Highly volatile sub-
stances are placed in combustion capsules (gelatin capsules ore acetobutyrate cap-
sules, see Accessories) and are burned together with the capsules.
The igniters (cotton thread) must be completely burned as well. If unburned remain-
ders of the igniter are left over, the experiment must be repeated or a correction
must be introduced into the result through the extraneous energy.
Substances with high halogen content can cause corrosion to appear on the de-
composition vessel. Decomposition vessel C 5012 should be used for these pur-
poses.
&DOLEUDWLRQ
To ensure exact reproducible measurement results, the calorimeter system is cali-
brated after it is first placed in service, after maintenance work, after parts are re-
placed and at specific time intervals. During calibration, the heat capacity of the
calorimeter system is re-determined.
5HJXODU FDOLEUDWLRQ LV DEVROXWHO\ HVVHQWLDO WR PDLQWDLQ DFFXUDF\ RI PHDVXUH
PHQW )XUWKHUPRUH WKH V\VWHP PXVW EH FDOLEUDWHG LQ WKH RSHUDWLQJ PRGH WKDW
ZLOO EH XVHG IRU WKH H[SHULPHQW DGLDEDWLF LVRSHULEROLF RU G\QDPLF
For this purpose, a specific quantity of a reference substance is burned in the de-
composition vessel under the conditions of the experiment. Since the gross calorific
value of the reference substance is known, it is possible to use the increase in tem-
perature of the calorimeter system when the reference substance is burned to cal-
culate the heat capacity.
The reference substance for calorimetry that is recognized at an international level is
benzoic acid obtained from the National Bureau of Standards (NBS-Standard Sam-
ple 39), with a guaranteed gross calorific value.
,I D FDORULPHWHU LV EHLQJ RSHUDWHG ZLWK PRUH WKDQ RQH GHFRPSRVLWLRQ YHVVHO
WKH KHDW FDSDFLW\ RI WKH V\VWHP PXVW EH GHWHUPLQHG IRU HDFK GHFRPSRVLWLRQ
YHVVHO
For more detailed information on calibration, please refer to the appropriate stan-
dards as they are listed in Section 4 “Features of the system”.
/LTXLG DQG
KLJKO\ YRODWLOH
VXEVWDQFHV
+DORJHQV
☞
☞