beautypg.com

Loop supply overloaded, Empty empty – Ashcroft ATE-2 - Handheld Calibrator User Manual

Page 75

background image

73

UNIT WILL NOT DELIVER 24Vdc LOOP SUPPLY OUTPUT

If the HHC is an Intrinsically Safe base unit , then loop supply is

not available.

Reference the manual section 14k to enable and disable the

power supply. Attempt to enable the supply. Ensure that the

“24V” icon appears on the display of the HHC.

If the power supply has been overloaded and a warning message

shows on the display (supply current is specified at 35ma and will

overload at approximately 54-55ma)

Remove the overload from test circuit and enable the 24 VDC

power supply via MENU commands. (See section 14k)

Ensure that the supply is being measured between the”24Vdc

Out” jack and the “COM” output jack at the top of the HHC.

Check all connections and wires in the set up to ensure continuity

if not working;

Remove all system wires connected to the HHC jacks at the top

of the unit and place a volt meter directly across the “24Vdc Out”

(+) and the “COM” (-) jacks, and measure the output. If the icon is

on the display and the 24Vdc output is there (24V ± .5 volts) then

the HHC is operational and the system set up is faulty. Check the

set up. If the loop supply output is still non-operational consult the

factory.

DATALOG WILL NOT EXPORT TO THE SD CARD

Ensure that the internal datalog memory has a datalog loaded

into it (see section 15f vi. Review Datalogs). If there are no stored

datalogs, the HHC will display an error message .

Verify that SD card write protect lock is off.

Ensure that the SD card is fully inserted into the onboard

connector of the HHC.

The SD card may be faulty, change the SD card

In order to ensure compatibility with the majority of manufacturers, the

PTE-2 was developed using a number of different types of SD cards

try using one of these recommended SD cards. (See section 16g)

Refer to the manual and repeat steps required to export data.

If the HHC still displays an error message, consult the factory.

SECTION 16 b.

TROUBLE SHOOTING CONT.

0.11

inH2O

PSi

15.29

0.033

mH

Loop supply overloaded

empty empty