The Damping Function stabilizes a measured pressure signal.

When damping is activated, the value displayed, saved to a

Datalog or transmitted to a journal (eg. To HyperTerminal) is a

filtered value. The filtering is programmable to 16 incremental

levels, allowing the damping function to be tailored to the spe-

cific requirements of most any application. The default damping

level is 1. As the damping level is increased, additional electronic

filtering is applied and the response time is slowed. Values are

displayed on the HHC screen, and recorded by Datalogging at

the fastest update rate between 100 and 300 milliseconds, based

on how many inputs are active. As damping level is increased the

update time increases approximately 100 milliseconds per damp-

ing level. With a single pressure module installed and damping

set to maximum level 16 the display output can lag by up to 1.6

seconds from the actual process pressure. If two modules are

installed and or voltage/current monitoring is active the system

display and Datalog values lag will also increase. The longest

display lag with two pressure modules programmed for maximum

damping and the voltage/current monitor active is 4.8 seconds.
Set up of Damping Function
When in Measurement Mode:
1. Press DAMP key.
2. Press PORT SELECT key to choose Channel 1, Channel 2, or

both modules.

3. Press RIGHT ARROW and LEFT ARROW to move cursor to

desired damping level (16 possible positions).

4. Press ENTER to save the value and activate damping.
Once established in the set up MENU the damping process may

be enabled or disabled through the use of the DAMP key on the

keypad. To activate/deactivate the damping function:
1. Press the DAMP key on the keyboard.
2. Press PORT SELECT key to choose Channel 1, Channel 2, or

both modules.

3. Press ENTER to activate damping. Damping level will be

indicated below units of measure.

The HHC has two Datalogging modes, Manual and Auto-

matic. Manual mode is the default. Automatic Datalog mode is

described in MENU programmable features section 15f.
The first time the DATA STORE key is pressed following power

up of the HHC, a new Datalog file will be created. The Datalog

column headers will be labeled with date, user programmable

module description (default values are Channel 1, Channel 2,

Ext.) and the engineering unit of measure. Programming channel

and tag names are covered in the MENU driven Datalogging

set up section 15f. The first row of the Datalog will be populated

with measurements as shown on the LCD display. Subsequent

presses of the DATA STORE key will create manual data records

into this initial Datalog.

SECTION 14 f.

DAMPING FUNCTION

–0.12

in H 2 0

-0.012

PSi

<----------|----->

SECTION 14 g.

MANUAL DATALOG FUNCTION