beautypg.com

Ashcroft ATE-2 - Handheld Calibrator User Manual

Page 66

background image

64

9. Select Volts Engineering Units

a. Press MENU key
b. Press DOWN or UP ARROW key to highlight “Units…”
c. Press ENTER
d. Press DOWN or UP ARROW key to highlight “Select Units”
e. Press ENTER
f. Press DOWN or UP ARROW key to highlight “Channel 1

Module”, “

g. Press ENTER
h. Press DOWN or UP ARROW key to highlight volts
i. Press ENTER

10.

Voltage displayed on Channel 1 is the compensated 0-1 volt

signal. Record this “As Found/Left” value.

11.

Verify that displayed voltage is within ±50uV (0.00005 Volt) of

the voltage standard

12.

Increase the input voltage setting from the voltage standard

at desired intervals until reaching 1.00000 volts.

13.

Note input standard voltages and HHC displayed voltages

and calculate error.

14.

It is recommended to at minimum verify display voltage at

0.00000 and 1.00000 volts input.

15.

When test is complete, set voltage standard to 0.000 volts

and disconnect the leads from the HHC.

16.

To Exit voltage display mode, remove the Calibration Module

and install a pressure or temperature measurement module

into either bay or by cycling power to the HHC

17. This completes the “As found” or “As Left” Calibration check

HHC Base Unit Field Calibration Procedure

Base unit recalibration is a firmware based, menu driven pro-

cedure that provides prompts on the HHC display directing the

operator to apply precision voltage and current signals to desig-

nated terminals. Recalibration is accessed via the menu system.
Flip Screen Function is not applicable in Calibration Mode.
1.

Remove any pressure or temperature modules from both

system module bays.

2.

Insert the Calibration Module (312D224-01) into the Channel

1 module bay.

3.

Leave Channel 2 bay empty or insert a system protection

module.

4.

Turn on the precision voltage standard and allow it to warm

up to its stated accuracy

5. Set the voltage standard to a range to generate 0.00000v
6.

Connect the negative lead of the voltage source to the black

terminal on the Calibration Module

7.

Connect the positive lead of the voltage source to the red

terminal on the Calibration Module

8.

Power on the HHC using either USB or battery power and

allow it warm up to its specified accuracy (5 minutes)

9. Enter the Calibration Process via the Setup menu:

a. Press MENU key
b. Press DOWN or UP ARROW key to highlight “Setup…”

c. Press ENTER

d. Press DOWN or UP ARROW key to highlight “Calibrate”
e. Press ENTER

Cal Password screen appears. Enter Calibration Password

SECTION 15 n.

CALIBRATE BASE UNIT CONT.