Send identification command (ai), Start measurement command (am) – Vaisala WXT520 User Manual
Page 85

Chapter 7 __________________________________________________ Getting the Data Messages
VAISALA_______________________________________________________________________ 83
Send Identification Command (aI)
This command is used to query the device for the SDI-12 compatibility
level, model number, and firmware version and serial number.
Command format: aI!
The response:
a13ccccccccmmmmmmvvvxxxxxxxx
Example:
0I!013VAISALA_WXT520103Y2630000
Start Measurement Command (aM)
This command asks the device to make a measurement. The measured
data are not sent automatically and should be requested with a separate
Send data command aD.
where
a
=
Device address
I
=
Send identification command
!
=
Command terminator
where
a
=
Device address
13
=
The SDI-12 version number, indicating SDI-12
version compatibility; for example, version 1.3 is
encoded as 13
cccccccc
=
8-character vendor identification Vaisala_
mmmmmm =
6 characters specifying the sensor model number
vvv
=
3 characters specifying the firmware version
xxxxxxxx
=
8-character serial number
=
Response terminator