Polling with crc – Vaisala WXT520 User Manual
Page 79

Chapter 7 __________________________________________________ Getting the Data Messages
VAISALA_______________________________________________________________________ 77
Example of the response (the parameters included can be chosen
from the full parameter set of the commands aR1, aR2, aR3 and
aR5):
0R0,Dx=005D,Sx=2.8M,Ta=23.0C,Ua=30.0P,Pa=1028.2H,
Rc=0.00M,Rd=10s,Th=23.6C
For selecting the parameter set in the response message, see
Sensor and Data Message Settings, on page 107
Polling with CRC
Use the same data query commands as in the previous sections but type
the first letter of the command in lower case and add a correct three-
character CRC before the command terminator. The response contains
also a CRC. For more information about the CRC-computation see
Appendix C, CRC-16 Computation, on page 161
.
Requesting a wind data message with a CRC:
Command format: ar1xxx
Example of the response (the parameter set is configurable):
0r1,Dn=236D,Dm=283D,Dx=031D,Sn=0.0M,Sm=1.0M,Sx=2.2MLFj
where the three characters before
response.
where
a
=
Device address
r1
=
Wind message query command
xxx
=
Three-character CRC for ar1 command
=
Command terminator
NOTE
The correct CRC for each command can be requested by typing the
command with an arbitrary three-character CRC.