Table 11, Error messaging/text messages table, Eir interpretation are shown in – Vaisala WXT520 User Manual
Page 138

User’s Guide ______________________________________________________________________
136 _________________________________________________________________ M210906EN-C
0xUabc!0TX,Use chksum CCb
the 0xU command)
Chapter 10, Troubleshooting, on page 133
error, please contact Vaisala Service Center, see
.
Table 11
Error Messaging/Text Messages Table
Text message
identifier (in
NMEA 0183 v3.0
protocol only)
Text Message
Interpretation and action
01
Unable to measure error The requested parameters are not activated in the
message, see
Chapter 8, Sensor and Data Message
, and check the parameter
selection fields.
02
Sync/address error
The device address in the beginning of the command is
invalid. Ask the device address with the ?! (SDI-12) or
?
command with the correct address.
03
Unknown cmd error
The command is not supported, use the correct
command format, see
.
04
Profile reset
Checksum error in configuration settings during power-
up. Factory settings used instead.
05
Factory reset
Checksum error in calibration settings during power-up.
Factory settings used instead.
06
Version reset
New software version in use.
07
Start-up
Software reset. Program starts from the beginning.
08
Use chksum xxx
Given checksum not correct for the command. Use the
proposed checksum.
09
Measurement reset
The ongoing measurement of all the sensors interrupted
and started from the beginning.
10
Rain reset
The precipitation sensor counter reset.
11
Inty reset
Precipitation sensor intensity counter reset.
NOTE
When WXT520 is used on an RS-485 bus with other polled devices,
the error messaging feature shall always be disabled. This is done by
the following command: 0SU,S=N