Specifications, 70 conductivity default settings – Sper Scientific 860031 pH - mV Bench-Top Meter User Manual

Page 70

background image

70

Conductivity Default Settings

Program Function

Default Display Note

P1.0
P1.1

Memory transmitting
MEM sent by RS232

No default “tr”

“out”

Follow Cond
or TDS of
Normal Mode

P2.0
P2.1

MEM clear
CLR confirm


Always
defaults
“no”

“CLr”
“no” or
“yes”

Follow Cond
or TDS of
Normal Mode

P3.0
P3.1~3.5

CAL view
Cal solution value


14.13 uS,
141.3 uS,
1413 uS,
14.13 mS,
141.3 mS

“CAL”
Cond/
TDS/
SALT
solution
value


Ra1~Ra5

P4.0
P4.1~4.5

CELL
Constant


1.000

“CELL”

Ra1~Ra5

P5.0
P5.1
P5.2
P5.3
P5.4

Temp setting
ATC/MTC
Tc
M Temp
TDS Factor


ATC
2.1%
25°C
0.500

“COEF”
“Auto” or
“NAn”



0.0% ~ 10.0%
0.300 ~ 1.000

P6.0
P6.1

Ready function
Enable or disable


“yes”

“rdy”
“no” or
“yes”

P7.0
P7.1

Temp unit
Select °C or °F


°C

“U”
“C” or “F”

P8.0
P8.1~8.6

Real time clock
Setting YMD, HMS


No default

“rtc”
“rtc”

P9.0

P9.1

RESET

Reset confirm



Always
defaults
“no”

“rSt”

“no” or
“yes”

Cond/TDS/
SALT
reset