Specifications, 70 conductivity default settings – Sper Scientific 860031 pH - mV Bench-Top Meter User Manual
Page 70
![background image](/manuals/404659/70/background.png)
70
Conductivity Default Settings
Program Function
Default Display Note
P1.0
P1.1
Memory transmitting
MEM sent by RS232
No default “tr”
“out”
Follow Cond
or TDS of
Normal Mode
P2.0
P2.1
MEM clear
CLR confirm
Always
defaults
“no”
“CLr”
“no” or
“yes”
Follow Cond
or TDS of
Normal Mode
P3.0
P3.1~3.5
CAL view
Cal solution value
14.13 uS,
141.3 uS,
1413 uS,
14.13 mS,
141.3 mS
“CAL”
Cond/
TDS/
SALT
solution
value
Ra1~Ra5
P4.0
P4.1~4.5
CELL
Constant
1.000
“CELL”
Ra1~Ra5
P5.0
P5.1
P5.2
P5.3
P5.4
Temp setting
ATC/MTC
Tc
M Temp
TDS Factor
ATC
2.1%
25°C
0.500
“COEF”
“Auto” or
“NAn”
0.0% ~ 10.0%
0.300 ~ 1.000
P6.0
P6.1
Ready function
Enable or disable
“yes”
“rdy”
“no” or
“yes”
P7.0
P7.1
Temp unit
Select °C or °F
°C
“U”
“C” or “F”
P8.0
P8.1~8.6
Real time clock
Setting YMD, HMS
No default
“rtc”
“rtc”
P9.0
P9.1
RESET
Reset confirm
Always
defaults
“no”
“rSt”
“no” or
“yes”
Cond/TDS/
SALT
reset
SPECIFICATIONS