ERROR CODES

E02

Reading is under the lower range limit; See

page 67 for range specifications for all

parameters

E03

Reading

is

over

the upper range limit; See

page 67 for range specifications for all

parameters

E04

Error in measuring original data (damaged

temperature sensor or temperature out of

specicifications) results in conductivity or pH

value error. E02 or E03 will also appear in the

temperature

column.

If E04 is caused by high

liquid temperature (E03), cool down the liquid

temperature.

E12

Factory

calibration

data error (pH); Restart

the

meter

E13

Slope or Offset value of the pH probe is out of

range

E16

Factory calibration data error (conductivity);

Restart

the

meter

E17

Cell constant of the conductivity probe is out

of range; Restart the meter

E31

Measuring circuit failure; Restart the meter

E32

Memory Integrated Circuit failure