APPENDICES

APPENDIX A:

CONDUCTIVITY to TDS CONVERSION FACTORS

442 stands for:
40% sodium sulfate, 40% sodium bicarbonate and 20%
sodium chloride.

APPENDIX B:

CALCULATING TDS CONVERSION FACTORS

The meter can be calibrated using TDS calibration standard
solutions. The calibration standard requires the TDS value
at a standard temperature such as 25°C. To determine the
conductivity-to-TDS conversion factor, use the following
formula:

Factor=Actual TDS ÷ Actual Conductivity @ 25°C

Definitions:

Actual TDS: Value from the solution bottle label or from a
standard buffer, which is made using high purity water and
precisely weighted salts.

TDS KCl

TDS NaCl

TDS 442

ppm

value

Factor ppm

value

Factor ppm

value

Factor

1413 uS
2070 uS
2764 uS
8974 uS

12,880 uS
15,000 uS

80 mS

744.7

1045
1382
5101
7447
8759

52,168

0.527

0.5048

0.5

0.5685
0.5782
0.5839
0.6521

702.1

1041

1414.8

4487
7230
8532

48,384

0.4969
0.5029
0.5119

0.5

0.5613
0.5688
0.6048

1000
1500

2062.7

7608

11,367
13,455
79,688

0.7078
0.7246
0.7463
0.8478
0.8825

0.897

0.9961

Conductivity

at 25°C