beautypg.com

Setup mode – Sper Scientific 860031 pH - mV Bench-Top Meter User Manual

Page 18

background image

18

SETUP MODE

View Slope & Offset (pH Probe)

1. Press MODE to select the probe type as pH.

2. Press SET for 2 seconds to enter setup.

3. Press until “ELE” appears in the middle display and

P3.0 appears in the lower display.

4. Press to enter P3.1, the LCD displays one of

four available slope values; P3.1, P3.2, P3.3, P3.4. If
the value is less than 75% or more than 115%, change
the probe immediately.

5. Press to enter P3.2, P3.3, and P3.4.

Note...
The solution range differs between NIST and Custom
buffers.