beautypg.com

Setup mode – Sper Scientific 860031 pH - mV Bench-Top Meter User Manual

Page 19

background image

19

SETUP MODE

6. Press to enter P3.5 and view the offset value. The

offset value is the mV value of pH 7 (default 0.0). The
offset value will be different after calibration. If the value
is outside the range of ± 60 mV, replace the probe.

Calibration Review (Conductivity Probe)

This feature allows you to review which range has been cal-
ibrated and the last calibration value. The program re-
views the probe calibration data of Conductivity, TDS or
SALT.

Note…
If the range is not yet calibrated, the LCD will display the
default value. There are 5 total calibration ranges for
Conductivity, TDS and SALT.

Range 1~3: Conductivity or TDS value
Range 4~5: Conductivity, TDS or SALT value

1. Press MODE to select the probe program.

2. Press SET for 2 seconds to enter setup.

3. Press to select CAL. “CAL” appears on the middle of

the LCD and P3.0 appears on the lower portion.