The procedure for salinity calibration is almost the same as
the procedure for conductivity calibration. Differences are
as follows:

1. In Measurement Mode, press MODE to select the mode

as SALT and press CAL to enter Calibration Mode.

2. There are two measuring ranges for salinity: 0 to 11.38

ppt and 0 to 80.0 ppt. Please select a NaCl standard
that is near the sample value you are measuring.