beautypg.com

Syntax, Relpresscal.a0[=n, Relpresscal.a1[=n – ROTRONIC MBW 373 User Manual

Page 46: Relpresscal.a2[=n, Relpresscal.a3[=n, Relpresscal.avg[=n, Relpresscal.avgband[=i, Relpresscal.senscode[=i, Relpresscal.coeftype[=i, Flow sensor calibration coefficients

background image

42

Relative Pressure Sensor Calibration Coefficients (if so equipped)

Syntax Function
RelPressCal.a0[=n][?]

Polynomial 0

th

order (Zero) coefficient

RelPressCal.a1[=n][?]

Polynomial 1

st

order (Span) coefficient

RelPressCal.a2[=n][?]

Polynomial 2

nd

order (Linearity) coefficient

RelPressCal.a3[=n][?]

Polynomial 3

rd

order (Linearity2) coefficient

RelPressCal.avg[=n][?]

Amount of averaging applied to measurement,
~100

RelPressCal.avgBand[=i][?]

Only average when measurements remain within
this band, ~ 1000 Pa

RelPressCal.sensCode[=i][?]

=21 for voltage output sensor

RelPressCal.coefType[=i][?]

=2 for polynomial type coefficients

SaveCal=373

Save these and other calibration parameters

Flow Sensor Calibration Coefficients

Syntax Function
FlowCal.a0[=n][?]

Polynomial 0

th

order (Zero) coefficient

FlowCal.a1[=n][?]

Polynomial 1

st

order (Span) coefficient

FlowCal.a2[=n][?]

Polynomial 2

nd

order (Linearity) coefficient

FlowCal.a3[=n][?]

Polynomial 3

rd

order (Linearity2) coefficient

FlowCal.avg[=n][?]

Amount of averaging applied to measurement, ~0

FlowCal.avgBand[=i][?]

Only average when measurements remain within
this band, ~ 0 l/m

FlowCal.sensCode[=i][?]

=21 for voltage output sensor

FlowCal.coefType[=i][?]

=2 for polynomial type coefficients

SaveCal=373

Save these and other calibration parameters

System Identification

Syntax Function
ID?

Returns a string containing instrument
identification, ie. DPM 373LX

IDN?

Returns only numeric portion of identifier, ie. 373

Saving Changes

Syntax Function
SaveCfg=373

Save configuration parameters

SaveCal=373

Save calibration parameters

SavePID=373

Save PID parameters