Syntax, Inttubetempcal.a0[=n, Inttubetempcal.a1[=n – ROTRONIC MBW 373 User Manual
Page 45: Inttubetempcal.a2[=n, Inttubetempcal.a3[=n, Inttubetempcal.avg[=n, Inttubetempcal.avgband[=i, Inttubetempcal.senscode[=i, Inttubetempcal.coeftype[=i, External tube temperature calibration coefficients
41
Internal Tube Temperature Calibration Coefficients (if so equipped)
Syntax Function
IntTubeTempCal.a0[=n][?]
Polynomial 0
th
order (Zero) coefficient, ~ 0
IntTubeTempCal.a1[=n][?]
Polynomial 1
st
order (Span) coefficient, ~ 0.01
IntTubeTempCal.a2[=n][?]
Polynomial 2
nd
order (Linearity) coefficient, ~0
IntTubeTempCal.a3[=n][?]
Polynomial 3
rd
order (Linearity2) coefficient, ~ 0
IntTubeTempCal.avg[=n][?]
Amount of averaging applied to measurement, ~40
IntTubeTempCal.avgBand[=i][?]
Only average when measurements remain within
this band, ~ 1°C
IntTubeTempCal.sensCode[=i][?]
=37 for 10K thermistor
IntTubeTempCal.coefType[=i][?]
=2 for polynomial type coefficients
SaveCal=373
Save these and other calibration parameters
External Tube Temperature Calibration Coefficients
Syntax Function
ExtTubeTempCal.r0[=n][?]
Callendar VanDusen R0 coefficient, ~ 100.00
ExtTubeTempCal.a[=n][?]
Callendar VanDusen A coefficient, ~ 3.9083e-3
ExtTubeTempCal.b[=n][?]
Callendar VanDusen B coefficient, ~ –5.775e-7
ExtTubeTempCal.c[=n][?]
Callendar VanDusen C coefficient, ~ -4.183e-12
ExtTubeTempCal.avg[=n][?]
Amount of averaging applied to measurement, ~40
ExtTubeTempCal.avgBand[=i][?]
Only average when measurements remain within
this band, ~ 1°C
ExtTubeTempCal.sensCode[=i][?]
=42 for 385 RTD
ExtTubeTempCal.coefType[=i][?]
=3 for Callendar VanDusen type coefficients
SaveCal=373
Save these and other calibration parameters
Pre Cooler Temperature Calibration Coefficients
Syntax Function
PreTempCal.r0[=n][?]
Callendar VanDusen R0 coefficient, ~ 100.00
PreTempCal.a[=n][?]
Callendar VanDusen A coefficient, ~ 3.9083e-3
PreTempCal.b[=n][?]
Callendar VanDusen B coefficient, ~ –5.775e-7
PreTempCal.c[=n][?]
Callendar VanDusen C coefficient, ~ -4.183e-12
PreTempCal.avg[=n][?]
Amount of averaging applied to measurement, ~40
PreTempCal.avgBand[=i][?]
Only average when measurements remain within
this band, ~ 1°C
PreTempCal.sensCode[=i][?]
=42 for 385 RTD
PreTempCal.coefType[=i][?]
=3 for Callendar VanDusen type coefficients
SaveCal=373
Save these and other calibration parameters
Absolute Pressure Sensor Calibration Coefficients
Syntax Function
AbsPressCal.a0[=n][?]
Polynomial 0
th
order (Zero) coefficient
AbsPressCal.a1[=n][?]
Polynomial 1
st
order (Span) coefficient
AbsPressCal.a2[=n][?]
Polynomial 2
nd
order (Linearity) coefficient
AbsPressCal.a3[=n][?]
Polynomial 3
rd
order (Linearity2) coefficient
AbsPressCal.avg[=n][?]
Amount of averaging applied to measurement, ~40
AbsPressCal.avgBand[=i][?]
Only average when measurements remain within
this band, ~ 1000 Pa
AbsPressCal.sensCode[=i][?]
=21 for voltage output sensor
AbsPressCal.coefType[=i][?]
=2 for polynomial type coefficients
SaveCal=373