beautypg.com

FUTEK IPM52x (D520) Digital Display User Manual

Page 32

background image

31

Configuration of 9600 baud, NO Parity Check, 8 DATA Bits,

with CTS handshaking available.

SECTION 5 SPECIFICATIONS:

±±±±20 mA

+/- 30 mV +/- 100 mV +/- 200 mV +/- 2V

±±±±10V

Accuracy 0.02%

0.02%

0.02%

0.02%

0.02%

0.02%

Zero Drift Autozero Autozero Autozero Autozero

Autozero Autozero

Span Drift 10 ppm/°C 10 ppm/°C 10 ppm/°C 10 ppm/°C 10 ppm/°C 10 ppm/°C
Input

Impedance

50 Ohms

10 Mohm

10 Mohm

10 Mohm

10 Mohm

10 Mohm

Resolution 1 uA

1 uV

10 uV

10 uV

100 uV

1 mV

Over-range

Capability,

minimum

10%

10%

10%

10%

10%

10%

Normal Mode Rejection

Ratio

63 bB, 50/60 Hz

Common Mode

Rejection Ratio

130 dB, 50/60 Hz

Response Time

100 msec max to display (10 updates / sec)

Internal Resolution

22 bits

Conversion Time

100 mS

Digital filter

10 Hz

Warm Up Time

15 minutes

Operating Temp range -20°C to 60°C
Storage Temp range

-20°C to 85°C

Humidity

To 95%, non-condensing

AC supply voltage

115 or 230 VAC

Power Consumption

15 Watts maximum (all options installed)

Isolation

Signal Input to Earth Ground: Safety rated to 450 Vrms, 2200Vp high voltage test
Option board to Earth Ground: Safety rated to 450 Vrms, 2200Vp high voltage test
Option board to Signal Input: 400 Vp
Power to Earth Ground:

Safety rated to 500 Vrms, 4000 Vp high voltage test

Shunt Cal. Relay

Contact Rating : 30VDC @ 1 A