13
•
“CHAN1”, (Only if CHAN is not disabled from above),
is flashed with current selected number. Valid values 0-
15.
0
Don’t disable any channel 1
1
Skip input range selection "INPUT"
2
Skip user calibration branch "USER"
4
Skip input calibration branch "INPUT"
8
Skip shunt calibration branch "SHUNT"
•
“CH” (Only if CHAN is not disabled from above) is
flashed with current selected number. Valid values 0-7.
0
Don’t skip Channel 2
1
Skip input range selection "INPUT"
2
Skip user calibration branch "USER"
4
Skip input calibration branch "INPUT"
•
“CH3” (Only if CHAN is not disabled from above) is
flashed with current selected number. Valid values 0-1.
0 Don’t skip Channel 3
1
Skip Channel 3 "CHAN 3"
•
“OUTPUT” (Only if OUTPUT is not disabled from
above) is flashed with current selected number. Valid
values 0-31.
0 Don’t disable any output menus
1
Disable Alarm 1 menu
2
Disable Alarm 2 menu
4
Disable Alarm 3 menu
8
Disable Alarm 4 menu
16
Disable the Source menu