20

c.

Channel 1 only has a third method of setup

"SHUNT" cal method (Also known as R-CAL):

1.

Raw input is displayed. This can be used to make sure

the correct input is being applied and/or that the meter is

properly measuring it. Push

ENTER to advance.

2.

"DEC PT" selects decimal point. Push

SELECT key for

displaying decimal point selections and

ENTER key to

confirm choice.

3.

"DSP LO" (display low). Enter in display value

corresponding to low input value.

4.

"DSP Hi" (display high). enter in display value

corresponding to shunt calibration resistor.

5.

At this time the unit stores the values and waits for a

Function key (F1-F4) programmed for shunt calibration

to start the following calibration.

a.

DPM measures input & sets display low equal to

input.

b.

Shunt cal relay is closed automatically.

c.

5 sec later DPM measures input & sets display

high equal to shunt cal input.

d.

Shunt relay opened.

e.

Return to the RUN MODE.

2.

Channel 3 (Math channel)

a.

"TYPE” Select for Channel 3 to be one of following:

"CHAN 1"

Channel 1’s raw input count

"CHAN 2"

Channel 2’s raw input count

"Add"

Sum of Channel 1 + 2

"Sub"

Difference of Channel 1 – 2

“Div”

Result of Channel 1

÷ 2

“Mul”

Multiply of CH1 x CH2

“P1-P2” Peak1 – Peak2

b.

Raw input is displayed. For exact calibration purposes you can

input the low and high inputs and record their value for entry in

the low and high input values below.

NOTE: The low and high