IV. Specifications

CONTROL UNIT

PARAMETER DETAIL

SPECIFICATION

NOTE

Input Voltage

115V~

97–132V~, 47–63Hz

Fuse: 0.032A, Slo-Blo 250V, Type ‘T’

Fuse: 0.050A, Slo-Blo 250V, Type ‘T’ when Self

Check Option is installed

Selector Switch to 115V

230V~

195–264V~, 47–63Hz

Fuse: 0.032A, Slo-Blo 250V, Type ‘T’

Fuse: 0.050A, Slo-Blo 250V, Type ‘T’ when Self

Check Option is installed

Selector Switch to 230V

24VDC

22–26VDC

Fuse: 0.250A, 250V, Type ‘T’

No Selector Switch

Optional Model

Input Power

6 Watts Max.

Measurement Units

picoamperes (pA)

1 x 10 ⎯ ¹² Amp

Standard

5.0pA – 5000pA

Detection Level/Range

Optional

0.5pA – 5000pA

Electronic Accuracy

± 5% of Range, std

± 1% of Range, optional

Over Full Temp Range

Operating

-13

°F to +160°F (-25°C to +70°C)

Temperature Range

Storage

-40

°F to +185°F (-40°C to +85°C)

Type Form

A

(SPST)

Relay Outputs

Rating

5A @ 240V~ Resistive **Customer must provide a

8A (maximum) fuse in series with relay load.

Type Isolated

Rating 470Ω Max. Loop Impedance

4-20mA Output (Optional)

Span

Adjustable Via Keypad

Default: 1000pA LOG

Type

Isolated, Multi-drop, ½ Duplex

RS-485 Network (Optional)

Settings

19,200 BPS, 8 Data Bits, 1 Stop Bit, No Parity

Modbus/RTU Protocol

Enclosure Type

Painted Cast Aluminum NEMA 4X or optional CSA

Approved Enclosure

Other optional

PARTICULATE SENSOR

NPT

½” NPT Thread

QC

1.5” Quick-Clamp (Mounting ferrule is 1” NPT)

FL

ANSI Flange (2”, 150# is typical)

Mounting

Other Others

Available

Other optional

Probe

Stainless Steel

With Protective Teflon Layer

Materials

Nipple/Mount Stainless

Steel

-40

°F to +250°F (-40°C to +120°C)

-40

°F to +450°F (-40°C to +232°C)

-40

°F to +800°F (-40°C to +426°C)

Process Temperature Range

Operating

> 800

°F

Higher Optional

Ambient Temperature Range

Operating

-40

°F to +160°F Maximum (-40°C to +70°C)

Maximum Must be Calculated for each application

T-Code Ratings for

AHZ1 option only

Pressure Range

Operating

Full Vacuum to 10PSI (2.11kg/cm²)

Higher Optional

Type

Painted Cast Aluminum

Other - Consult Factory

Enclosure

Rating

NEMA 4X (IP 66)

Other - Consult Factory