Particulate Monitoring Systems

Installation & Operating Manual

Document No. 210-1003-M

Page 17

©2007

6.5 Long Term Averaging

The long term averaging function provides a rolling average of the real-time readings over time.































Averaging setpoints are accessed through averaging setup screen. To enter the averaging setup screen,
press the SETUP key while viewing the MAIN AVERAGE screen. Navigation within the setup
menu is similar to that described in the previous setup screens.

Press the SETUP key from the MAIN AVERAGE screen to enter the

AVERAGE setup screen. Use the UP/DOWN arrow keys to adjust

settings to the desired values. Press the ENTER key to save a value and

proceed to next setup screen. Press the SETUP/ESCAPE key to return

to the main average screen without saving the current parameter’s value.

Menu Item

Range

Definition

Average

Period

0-360

Min

Number of minutes the real-time

reading is averaged to compute the

average reading.

Clear

Average

YES/NO

When YES is selected and the enter

key is pressed the averaging is cleared

and the average reading is initialized

to 0.

Press the ENTER key to toggle between the Real-Time display and the Average display.

Main

Screen

Units

Shown

Description

Real-Time

Display

pA

Displays the real-time reading. Before display, the real-time

reading is processed through a low-pass input signal filter

which smoothes the reading. See the low-pass “Signal Filter

Frequency” setup in the Menu 3 setup section.

Average

Display

ApA

Displays a selectable rolling time average of the real-time

reading. The average period (in minutes) is adjustable

through the average setup screen. The word “AVG” is quickly

shown in place of the readout once every five seconds,

indicating that the displayed and output reading is an

average.