beautypg.com

Iii. approvals and certifications – Dwyer DPM User Manual

Page 4

background image

Particulate Monitoring Systems

Installation & Operating Manual

Document No. 209-1025-D

iv

©2007

III. Approvals and Certifications

CE Conformant

The Control Units and Particulate Sensors conform to the appropriate country standards and
governing regulations listed below:

• EN 61010/1993 “Safety requirements for electrical equipment for measurement, control
and laboratory use.”

• EN

55011/1991 “Limits and methods of measurement of radio interference

characteristics of industrial, scientific and medical equipment”. Class A: Industrial and
commercial.

• EN50082-1/1993 “Electromagnetic compatibility – Generic immunity standard”. Part 1:
Residential, commercial and light industry.

CSA Certified

This Particulate Monitoring system is certified by the Canadian Standards Association (to
US and Canadian Standards) for use in hazardous locations as specified below:

Particulate Sensor:

Intrinsically Safe for Hazardous Locations

PMS-AHZ1 Intrinsically Safe Particulate Sensor in the following areas:

Class I, Division 1, Groups A, B, C, and D

Class II, Division 1, Groups E, F, G

Class III

For use with the Control Units listed below, 70ºC maximum ambient.

Note: There is no temperature rise caused by electrical components. Temperature
code is based only on ambient temperature, (e.g. a 200°C process requires a T3
rating.)

Control Unit:

For use in Ordinary Locations Only

DPM-AHZ1 Control Unit rated 115 Vac/230 Vac, 0.1A or 24Vdc, 0.25A, -25ºC to
70°C, with intrinsically-safe output to Particulate Sensor.

DPM-AHZ1 Control Unit - Approved for use with PMS-AHZ1
sensors, where the PMS-AHZ1 sensor is the only component located in
the hazardous area.

This manual is related to the following products: