Iii. approvals and certifications – Dwyer DPM User Manual
Page 4
![background image](/manuals/372023/4/background.png)
Particulate Monitoring Systems
Installation & Operating Manual
Document No. 209-1025-D
iv
©2007
III. Approvals and Certifications
CE Conformant
The Control Units and Particulate Sensors conform to the appropriate country standards and
governing regulations listed below:
• EN 61010/1993 “Safety requirements for electrical equipment for measurement, control
and laboratory use.”
• EN
55011/1991 “Limits and methods of measurement of radio interference
characteristics of industrial, scientific and medical equipment”. Class A: Industrial and
commercial.
• EN50082-1/1993 “Electromagnetic compatibility – Generic immunity standard”. Part 1:
Residential, commercial and light industry.
CSA Certified
This Particulate Monitoring system is certified by the Canadian Standards Association (to
US and Canadian Standards) for use in hazardous locations as specified below:
Particulate Sensor:
Intrinsically Safe for Hazardous Locations
PMS-AHZ1 Intrinsically Safe Particulate Sensor in the following areas:
Class I, Division 1, Groups A, B, C, and D
Class II, Division 1, Groups E, F, G
Class III
For use with the Control Units listed below, 70ºC maximum ambient.
Note: There is no temperature rise caused by electrical components. Temperature
code is based only on ambient temperature, (e.g. a 200°C process requires a T3
rating.)
Control Unit:
For use in Ordinary Locations Only
DPM-AHZ1 Control Unit rated 115 Vac/230 Vac, 0.1A or 24Vdc, 0.25A, -25ºC to
70°C, with intrinsically-safe output to Particulate Sensor.
DPM-AHZ1 Control Unit - Approved for use with PMS-AHZ1
sensors, where the PMS-AHZ1 sensor is the only component located in
the hazardous area.