Particulate Monitoring Systems

Installation & Operating Manual

Document No. 210-1003-M

Page 39

©2007

Service Class

Instance Attribute Data

0x45 0x67

0x89

0xAB 0xCD

6. BOOTP operation within the module must be DISABLED. To DISABLE BOOTP operation

the following explicit message must be sent to the module:

Service Class

Instance Attribute Data

0x10 0x64

0x01

0x6E 0x00

7. Cycle power to the module. Once power is re-applied the module will come online with the

new IP address and Subnet mask. BOOTP operation should be disabled at this point so no
BOOTP requests will be issued by the module.

8. A network ping should be used to verify that the IP address and Subnet mask have been set

correctly.


The control unit must always be configured with Modbus®/RTU RS-485 network address set to 1.

Input (T->O) Assembly Object (Class 0x04, Instance 0x65)

The following is a mapping of I/O values for the standard Ethernet/IP® assembly object instance
0x65. These are values that are sent from the Ethernet/IP® server (control unit) to the Ethernet/IP
client (typically a PLC). The input assembly size is 13 (16 bit words).

Data Types

Bool – 1 bit Boolean discrete data

Usint – 16 bit unsigned integer analog data

Sint – 16 bit signed analog data

Float – 32 bit signed floating point analog data

Floating point values are represented in IEEE format (32-bit) where the 1

st

register is the LOW

word and the 2

nd

register is the HIGH word.

Input (T->O) Assembly Object (Class 0x04, Instance 0x65) Size 13 (16 bit words)

16-Bit

Word

Modbus®

Range Range Units Data

Type

Description

Address Address Min Max

0

-

0

65535

Usint

Network status (0=ok, >0=error)

1

-

0

65535

Usint

Alarm status 1-16 (1 = alarm, 0 = ok)

1:0

03073

0

1

Bool

Alarm status – Max alarm #1 (1=ON, 0=OFF)

1:1

03074

0

1

Bool

Alarm status – Pre-Visible alarm #2 (1=ON, 0=OFF)

2

-

0

65535

Usint

Self check status information

2:0

03247

0

1

Bool

Zero check status (1=RUNNING, 0=NOT RUNNING)

2:1

03248

0

1

Bool

Span check status (1=RUNNING, 0=NOT RUNNING)

2:2

03249

0

1

Bool

Full system zero check status (1=RUNNING, 0=NOT
RUNNING)

2:3 - 0

1

Bool

Not

Used

2:4

03251

0

1

Bool

Zero check remote activate (1=ACTIVATED, 0=NOT
ACTIVATED)

2:5

03252

0

1

Bool

Span check remote activate (1=ACTIVATED, 0=NOT
ACTIVATED)

2:6 - 0

1

Bool

Not

Used

2:7

03254

0

1

Bool

Probe and cable check status (1=ERROR, 0=OK)

2:8

03255

0

1

Bool

Zero check status (1=ERROR, 0=OK)

2:9

03256

0

1

Bool

Span check status (1=ERROR, 0=OK)

2:10

03257

0

1

Bool

Full system zero check status (1=ERROR, 0=OK)