Er19 - Tried to run profile
1. Press the RUN/HOLD key, then press and release the
with 0 segments
SCROLL key until oFF or CtrL are displayed, then
press the DOWN key. This error occurs if a profile
number is selected in the Profile Continue mode for a
profile that has not been developed.
Er20 - Setpoint Error
1. Press the UP or DOWN key to change the setpoint to
a different value, then restore the needed setpoint.
2. Perfrom the CAL1 procedure as described in the
Calibration section (page 52).
Er21 - Pen Feedback
1. Perform the steps listed for pens not moving.
Error
Er36 - Incorrect Crystal
1. Turn off the power to the instrument, wait 5 seconds,
For Digital Comm.
then turn the power on.
Er37 - Incorrect Micro.
1. Turn off the power to the instrument wait 5 seconds,
For Digital Comm.
then turn the power on.
Er38 - Incorrect RAM
1. Turn off the power to the instrument, wait 5 seconds,
for profiler
then turn the power on.
2. Re-configure for non-profiler.
Momentary Er 70 -
1. May occur when incrementing or decrementing a
Controller unable to respond
value. Error clears itself.
within 250 milliseconds
Momentary Er 71 -
1. Computer program must wait longer for a response
Byte received before the
from the instrument.
response was transmitted
Momentary Er 72 -
1. Data received not valid, possible corruption on the
Incorrect Block Check
comm line. Possible noise problems. Check baud
character was received
rates, instrument addresses, line connections, and
termination.
Momentary Er 73 -
1. Check baud rate.
Byte received with incorrect
2. Possible noise problems. Check instrument
parity.
addresses, line connections, and termination.