beautypg.com

51 calibration 5.4, Calibration flow chart – West Control Solutions MRC 7000 Profiler Manual User Manual

Page 51

background image

51

Calibration 5.4

CAUTION: Do not attempt any calibrations without the proper test equipment that meets or
exceeds the specifications listed.

Press and release the SCROLL key until CAL appears on the display , then press the DOWN
key to enter the mode. The display will change to CAL1. Press the SCROLL key to advance
the display to the other calibration modes available. For two pen units, CAL2 and CAL3 will
only need to be required on TB4 to calibrate Pen 1 and Pen 2 inputs. Both TB 4 and TB 5
inputs must be calibrated for thermocouple inputs. Table 5-1, page 52, provides a listing of
field calibration routines. All instruments are calibrated prior to shipment from factory.

CALIBRATION FLOW CHART

ON
OFF

Actual Display

On/Off Display -
Use arrow keys
to turn on or off

Scroll Key

Numeric Display -
Use arrow keys
to change value

Up Arrow Key

Down Arrow

KEY

CAL2

CAL3

CAL4

CAL5

CAL6

CAL1

CAL

CAL7

CAL8

CAL9